4.7 Article

Ultra-trace determination of beryllium in occupational hygiene samples by ammonium bifluoride extraction and fluorescence detection using hydroxybenzoquinoline sulfonate

期刊

ANALYTICA CHIMICA ACTA
卷 584, 期 2, 页码 281-286

出版社

ELSEVIER
DOI: 10.1016/j.aca.2006.11.066

关键词

air monitoring; beryllium; extraction; fluorescence; trace analysis; workplace

向作者/读者索取更多资源

A highly sensitive molecular fluorescence method for measuring ultra-trace levels of beryllium has been previously described. The method entails extraction of beryllium workplace samples by 1% ammonium bifluoride (NH4HF2, aqueous), followed by fluorescence detection using hydroxybenzoquinoline sulfonate (HBQS). In this work, modification of the existing procedure resulted in a significant improvement in detection power, thereby enabling ultra-trace determination of beryllium in air filter and surface wipe samples. Such low detection limits may be necessary in view of expected decreases in applicable occupational exposure limits (OELs) for beryllium. Attributes of the modified NH4HF2 extraction/HBQS fluorescence method include method detection limits (MDLs) of < 0.8 ng to approximate to 2 ng Be per sample (depending on the fluorometer used), quantitative recoveries from beryllium oxide, a dynamic range of several orders of magnitude, and freedom from interferences. Other key advantages of the technique are field portability, relatively low cost, and high sample throughput. The method performance compares favorably with that of inductively coupled plasma-mass spectrometry (ICP-MS). (c) 2006 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据