4.5 Article

Inversion recovery prepared turbo spin echo sequences with reduced SAR using smooth transitions between pseudo steady states

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 57, 期 3, 页码 631-637

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.21170

关键词

inversion recovery; TRAPS; FSE; TSE; variable flip angles; STIR; FLAIR; hyperecho; SAR

向作者/读者索取更多资源

This study investigates the contrast behavior of 2D inversion recovery (IR) prepared turbo spin echo (TSE) sequences that use RF pulse schemes with variable low flip angles (hyperTSE) to reduce RIF power deposition. A framework of equations and calculations for adapting the sequence parameters is presented by which equivalent image contrast is achieved compared to conventional IR-TSE imaging. Although the inversion time (TI) and repetition time (TR) do not need to be changed, the echo time (TE) has to be prolonged such that the effective TE (TEeff) is preserved. Measurements in healthy volunteers confirmed this finding for IR-TSE sequences using different TIs: fluid attenuated inversion recovery (FLAIR), gray matter (GM)white matter (WM)-IR, and short tau IR (STIR). The results demonstrate that hyperTSE sequences enable high-quality IR-prepared imaging with a considerably reduced specific absorption rate (SAR).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据