Minimizing tissue-material interaction in microsensor for subcutaneous glucose monitoring

期刊

BIOSENSORS & BIOELECTRONICS
卷 22, 期 8, 页码 1625-1632

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2006.07.025

关键词

microsensor; rat's serum; subcutaneous; hematology; biocompatibility

向作者/读者索取更多资源

A new implantable electrocatalytic glucose sensor for subcutaneous glucose monitoring has been fabricated by immobilizing glucose oxidase on a chemically modified carbon fiber. The sensor was inserted subcutaneously on a male spraguely rat without any incision after dipping the microsensor in the rat's serum for 3 days. The so called stained microsensor, operated in the amperometric mode with an applied potential of +0.23 V versus Ag vertical bar AgCl, was able to directly measure the glucose concentration upon infusion of glucose. The results obtained were encouraging, with the response time was less than 2 s and the apparent Michaelis-Menten value at 5.1 +/- 0.5 mM. The stained microsensor shows good stability and reproducibility with constant response spanned over 25 days. Most common interferences in glucose analysis were minimized by the outerlayer Nafion((R)). Hematology examinations showed minimal material-tissue interaction. Use of such mechanical devices will allow a more refined understanding towards glucose control in diabetic patients as the implanted microsensor was not effected by biocompatibility failures. (c) 2006 Elsevier B.V. All rights reserved.