Temperature and SAR measurement errors in the evaluation of metallic linear structures heating during MRI using fluoroptic® probes

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 52, 期 6, 页码 1633-1646

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/52/6/006

关键词

-

向作者/读者索取更多资源

The purpose of this work is to evaluate the error associated with temperature and SAR measurements using fluoroptic (R) temperature probes on pacemaker ( PM) leads during magnetic resonance imaging (MRI). We performed temperature measurements on pacemaker leads, excited with a 25, 64, and 128 MHz current. The PM lead tip heating was measured with a fluoroptic (R) thermometer (Luxtron, Model 3100, USA). Different contact configurations between the pigmented portion of the temperature probe and the PM lead tip were investigated to find the contact position minimizing the temperature and SAR underestimation. A computer model was used to estimate the error made by fluoroptic (R) probes in temperature and SAR measurement. The transversal contact of the pigmented portion of the temperature probe and the PM lead tip minimizes the underestimation for temperature and SAR. This contact position also has the lowest temperature and SAR error. For other contact positions, the maximum temperature error can be as high as -45%, whereas the maximum SAR error can be as high as -54%. MRI heating evaluations with temperature probes should use a contact positionminimizing the maximum error, need to be accompanied by a thorough uncertainty budget and the temperature and SAR errors should be specified.