期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 848, 期 2, 页码 379-383出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2006.10.046
关键词
erlotimb; OSI-774; HPLC-MS/MS
An accurate and precise method was developed using HPLC-MS/MS to quantify erlotinib (OSI-774) and its O-desmethyl metabolite, OSI-420, in plasma. The advantages of this method include the use of a small sample volume, liquid-liquid extraction with high extraction efficiency and short chromatographic run times. The analytes were extracted from 100 mu L plasma volume using hexane:ethyl acetate after midazolam was added to the sample for internal standardization. The compounds were separated on a Phenomenex C-18 Luna analytical column with acetonitrile:5 mM ammonium acetate as the mobile phase. All compounds were monitored by tandem mass spectrometry with electrospray positive ionization. The intra-day accuracy and precision (% coefficient of variation, % CV) estimates for erlotinib at 10 ng/mL were 90% and 9%, respectively. The intra-day accuracy and precision estimates for OSI-420 at 5 ng/mL were 80% and 4%, respectively. This method was used to quantify erlotinib and OSI-420 in plasma of patients (n = 21) administered 150 mg erlotinib per day for non-small cell lung cancer. (c) 2006 Elsevier B.V. All rights reserved.