期刊
MICROCHIMICA ACTA
卷 158, 期 1-2, 页码 131-136出版社
SPRINGER WIEN
DOI: 10.1007/s00604-006-0691-x
关键词
acetaminophen; voltammetric determination; carbon nanotube; modified electrode
Single-wall carbon nanotubes (SWNT) were dispersed into water in the presence of dicetyl phosphate (DCP), and then a SWNT-DCP film-coated glassy carbon electrode ( GCE) was constructed. The electrochemical behavior of acetaminophen at bare GCE and SWNT-DCP modified GCE were compared, suggesting that the SWNT-DCP-modified GCE significantly enhances the oxidation peak current of acetaminophen. A sensitive and simple electrochemical method with a good linear relationship in the range of 1.0 x 10(-7) - 2.0 x 10(-5) mol L-1, was developed for the determination of acetaminophen. The detection limit is 4.0 x 10(-8) mol L-1 for 3-min accumulation. This method was successfully demonstrated with tablets.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐