4.7 Article

Detection of human chorionic gonadotrophin hormone using a label-free epoxysilane-modified capacitive immunosensor

期刊

APPLIED MICROBIOLOGY AND BIOTECHNOLOGY
卷 74, 期 6, 页码 1385-1391

出版社

SPRINGER
DOI: 10.1007/s00253-006-0778-7

关键词

capacitive immunoassay; human chorionic gonadotropin hormone

向作者/读者索取更多资源

A new and label-free capacitive immunosensor based on antibody- functionalized epoxysilane on a glassy carbon electrode has been developed for quantitative detection of human chorionic gonadotropin (hCG). Monitoring the changes in the capacitance signals of antibodies before and after the binding of the antigen provides the basis for an immunoassay. The performance and factors influencing the immunosensor were also studied. Under the optimized conditions, the developed immunosensor quantitatively detected serum hCG in the range of 18-450 mlU/ ml with a detection limit of 5.0 mlU/m1 (at 36). Thirty-five patients' sera were assayed by the proposed immunosensor, and the results agreed with those given by the commercial radioimmunoassay test kit, with correlation coefficient of 0.998. Further research about the intrinsic electroactivity of antibodies and their target molecules would surely provide new and sensitive screening assays as well as extensive data regarding their interaction mechanisms.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据