4.7 Article Proceedings Paper

New capability for generating and measuring small DC currents at NPL

期刊

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TIM.2007.890789

关键词

calibration; capacitance; current measurement; voltage measurement

向作者/读者索取更多资源

This paper describes two systems recently developed at the National Physical Laboratory (NPL), Teddington, U.K., for the generation and measurement of dc currents in the range 100 fA to 1 nA. Both systems are based on the capacitor charging technique. One system generates known currents, and constitutes a new reference at NPL. The other was designed to meet the traceability requirements of the Ionizing Radiation sections of NPL and combines commercially available instrumentation and standards to produce a reliable stand-alone system for measuring small currents.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据