Accuracy of two heterogeneity dose calculation algorithms for IMRT in treatment plans designed using an anthropomorphic thorax phantom

期刊

MEDICAL PHYSICS
卷 34, 期 5, 页码 1850-1857

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.2727789

关键词

-

资金

  1. NCI NIH HHS [CA081647, CA010953] Funding Source: Medline

向作者/读者索取更多资源

With the advent of intensity-modulated radiation therapy (IMRT), the inclusion of heterogeneity corrections is further complicated by the conformal delivery of many small beams forming steep dose gradients. Radiation treatment planning has evolved to take into account even small changes in tissue density so that the dose to tumor can be further optimized. However, different treatment planning systems incorporate different heterogeneity correction algorithms, and it is unclear whether any of these algorithms are superior to others in terms of accurately predicting delivered radiation doses relative to measurement in a clinical setting. The purpose of this study was to determine the accuracy of heterogeneity dose calculations from two widely used IMRT treatment planning systems (Pinnacle and Corvus) against measurement. These two systems handle heterogeneity dose corrections by means of a collapsed-cone convolution superposition algorithm and a finite-size pencil-beam algorithm with one-dimensional depth scaling correction, respectively. Treatment plans were generated by each system using an anthropomorphic thorax phantom, routine clinical lung tumor constraints, and a common prescribed dose. Dose measurements made by thermoluminescent detectors (TLDs) and radiochromic film positioned within the phantom's lung and offset tumor insert were then compared with the calculated values. The collapsed cone convolution superposition dose calculation algorithm provided clinically acceptable results (5% of the normalization dose or 3 mm distance to agreement) in the designed treatment plan and delivery. The pencil-beam algorithm with an effective pathlength correction showed reasonable agreement within the gross tumor volume, overestimated dose within a majority of the planning target volume, and underestimated the extent of the penumbral broadening, yielding only about 60% accuracy when judged by the above criterion. Even judged by a more generous criterion (+/- 7% / 7 mm), the results were clinically unfavorable (at only about 80% accuracy). To ascertain the dose in heterogeneous regions such as the tumor-lung interface and the peripheral lung dose near the tumor, the superposition convolution algorithm that accounts for lateral scatter and electron transport should be used. The use of the pencil-beam algorithm with only an effective pathlength correction may result in the dose to the target being overestimated. As a result, a full understanding of any treatment planning system's heterogeneity algorithm is required prior to clinical implementation. (c) 2007 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据