Validation of metal oxide semiconductor field effect transistor technology for organ dose assessment during CT: Comparison with thermoluminescent dosimetry

期刊

AMERICAN JOURNAL OF ROENTGENOLOGY
卷 188, 期 5, 页码 1332-1336

出版社

AMER ROENTGEN RAY SOC
DOI: 10.2214/AJR.06.0742

关键词

CT dosimetry; metal oxide semiconductor field effect transistor; radiation dose; radiologic physics; thermoluminescent dosimeter

向作者/读者索取更多资源

OBJECTIVE. The purposes of this study were to apply near-real-time dose-measurement technology with metal oxide semiconductor field effect transistors (MOSFETs) to the assessment of organ dose during CT and to validate the method in comparison with the thermoluminescent dosimeter (TLD) method. MATERIALS AND METHODS. Dosimetry measurements were performed in two ways, one with TLDs and the other with MOSFETs. Twenty organ locations were selected in an adult anthropomorphic female phantom. High-sensitivity MOSFET dosimeters were used. For the reference standard, TLDs were placed in the same organ locations as the MOSFETs. Both MOSFET and TLD detectors were calibrated with an X-ray beam equivalent in quality to that of a commercial CT scanner (half-value layer, approximate to 7 mm Al at 120 kVp). Organ dose was determined with a scan protocol for pulmonary embolus studies on a 4-MDCT scanner. RESULTS. Measurements for selected organ doses and the percentage difference for TLDs and MOSFETs, respectively, were as follows: thyroid (0.34 cGy, 0.31 cGy, - 8%), middle lobe of lung (2.4 cGy, 3.0 cGy, + 26%), bone marrow of thoracic spine (2.2 cGy, 2.5 cGy, + 11%), stomach (1.0 cGy, 0.93 cGy, - 6%), liver ( 2.5 cGy, 2.6 cGy, + 6%), and left breast ( 3.0 cGy, 2.9 cGy, - 1%). Bland-Altman analysis showed that the MOSFET results agreed with the TLD results ( bias, 0.042). CONCLUSION. We found good agreement between the results with the MOSFET and TLD methods. Near-real-time CT organ dose assessment not previously feasible with TLDs was achieved with MOSFETs. MOSFET technology can be used for protocol development in the rapidly changing MDCT scanner environment, in which organ dose data are extremely limited.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据