4.8 Article

Target delivery in a microfluidic immunosensor

期刊

BIOSENSORS & BIOELECTRONICS
卷 22, 期 11, 页码 2763-2767

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2006.12.017

关键词

mixer; grooves; microchannels; fluorescence; immunoassay; depletion

向作者/读者索取更多资源

A study is presented that examines the effect of rn icrofluidic mixing elements on direct and sandwich assays performed in microchannels. Patterned grooves were embossed in the top of microchannels made in PDMS using soft lithography. The grooves redirected the fluid flowing in the channel, enhancing delivery of the target from the bulk fluid to the surface and preventing the formation of a depletion layer at the surface. Comparing assays in grooved and plain channels demonstrated that the mixers improved assay results by 26-46%. A computational flow analysis showed that the grooves caused virtual particles in the bulk flow to come close to the surface (similar to 11 mu m) which is consistent with the signal increase seen experimentally. Direct assays for several concentrations of CY5-labeled biotin were performed in the microchannels. The mixers also improved signal intensity in sandwich assays for botulinum toxin which required mixing of the reagents as well as the direction of the target to the surface. (C) 2006 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据