4.7 Article

Simultaneous determination of copper and iron in automotive gasoline by X-ray fluorescence after pre-concentration on cellulose paper

期刊

TALANTA
卷 72, 期 3, 页码 1073-1076

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2006.12.042

关键词

gasoline; EDXRF; metals; iron; copper

向作者/读者索取更多资源

This paper proposes an alternative analytical method using energy dispersive X-ray fluorescence (EDXRF) to determine Fe and Cu in gasoline samples. In the proposed procedure, samples were distilled and the distillation residues were spotted on cellulose paper disk to form a uniform thin film and to produce a homogeneous and reproducible interface to the XRF instrument. The disks were dried at 60 degrees C for 20 min and copper and iron were determined directly in the solid phase at 6.40 and 8.04 keV, respectively. The calibration curves showed linear response in the 20-800 mu g L-1 concentration range of each metal. The precisions (repeatability) calculated from 15 consecutive measurements and defined as the coefficient of variation of solutions containing 100 mu g L-1 of Fe and Cu were 7.8 and 8.1%, respectively. The limits of detection (LOD), defined as the analyte concentration that gives a response equivalent to three times the standard deviation of the blank (n = 10), were found to be 10 and 15 mu g L-1 for Fe and Cu, respectively. The proposed method was applied to copper and iron determination in gasoline samples collected from different gas stations. (C) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据