4.5 Article

ASIC for small angle neutron scattering experiments at the SNS

期刊

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
卷 54, 期 3, 页码 541-548

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2007.893718

关键词

ADC; ASIC; gas detector; peak detector

向作者/读者索取更多资源

We present an ASIC for a He-3 gas detector to be used in small angle neutron scattering experiments at the Spallation Neutron Source in Oak Ridge. The ASIC is composed of 64 channels with low noise charge amplification, filtering, timing and amplitude measurement circuits, where an innovative current-mode peak-detector and digitizer (PDAD) is adopted. The proposed PDAD provides at the same time peak detection and A/D conversion in real time, at low power, and without requiring a clock signal. The channels share an efficient data sparsification and derandomization scheme, a 30-bit 256 deep FIFO, and low voltage differential signaling.