期刊
ANALYTICAL SCIENCES
卷 29, 期 2, 页码 271-274出版社
JAPAN SOC ANALYTICAL CHEMISTRY
DOI: 10.2116/analsci.29.271
关键词
-
资金
- Ministry of Education, Culture, Sports, Science and Technology, Japan
A method was developed for the determination of I-129 in soil samples that uses an ICP-MS equipped with an octopole reaction system. Oxygen was used as the reaction gas for reducing the background intensity of m/z 129, principally by Xe-129(+). The contribution of polyatomic ions, such as (IH2)-I-127+, could be effectively corrected for by assuming a production ratio ((IH2)-I-127+/I-127(+) = 3 x 10(-8)). The measured I-129/I-127 ratios in a NIST standard solution and Fukushima soil samples are consistent with the expected value within the analytical error. This method provides a powerful tool for investigating radioiodine contamination in Fukushima and elsewhere.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐