4.6 Article

A single-valve sequential injection manifold (SV-SIA) for automation of air-assisted liquid-phase microextraction: stopped flow spectrophotometric determination of chromium(VI)

期刊

ANALYTICAL METHODS
卷 5, 期 10, 页码 2497-2502

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c3ay40284c

关键词

-

资金

  1. Faculty of Science of P.J. Safarik University in Kosice [VVGS-PF-2012-25]

向作者/读者索取更多资源

A single-valve sequential injection setup is suggested for automation of an air-assisted liquid-phase microextraction. In order to eliminate parasitic signals, the stopped flow detection mode was employed. The aqueous phase (1.5 mL), containing chromium(VI), chloride ions (0.6 mol L-1), sulphuric acid (0.07 mol L-1) and dimethyl indocarbocyanine dye (1.33 x 10(-4) mol L-1), was dispensed into the extraction chamber and was followed by moderate injection of a 250 mu L organic phase and intense mixing by multiple air bubbling. The extraction phase was then submitted for stopped flow UV-Vis spectrophotometric detection at 560 nm wavelength. The calibration plot was linear up to 104 mu g L-1 of chromium(VI). The limit of detection, calculated as 3s of a blank test (n = 10), was found to be 4.5 mu g L-1. The developed procedure was applied for the determination of Cr(VI) in various water samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据