A novel micro tensile tester was developed to measure the viscoelastic properties of single isolated cells. A specimen cell was held with two glass micropipettes, the tips of which were coated with a urethane resin adhesive. One pipette was rigid and fixed in position, and the other was deflectable and could be moved by a laboratory-made piezoelectric actuator to stretch the specimen. The actuator was made of an originally designed lever mechanism which can magnify the deformation of conventional piezoceramic actuator and allowed for sufficient displacement up to 150 mu m. The distance between the two pipette tips was obtained by real-time image processing and kept constant following stepwise stretching with an accuracy of +/- 0.1 mu m by changing the voltage applied to the piezoceramics under feed-back control. The force applied to the cell was measured by the deflection of a cantilever part of the deflectable pipette. The stress relaxation test was performed for cultured rat aortic smooth muscle cells. The cells were stretched by 75% and their length was kept constant. Viscoelastic analysis with a standard linear solid showed that the relaxation time constant of the isolated smooth muscle cells was 250 +/- 42 s (n = 8, mean +/- S.E.M.) which was over six times longer than that reported for vascular endothelial cells and fibroblasts. (c) 2006 IPEM. Published by Elsevier Ltd. All tights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据