Cantilever spring constant calibration using laser Doppler vibrometry

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 78, 期 6, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.2743272

关键词

-

向作者/读者索取更多资源

Uncertainty in cantilever spring constants is a critical issue in atomic force microscopy (AFM) force measurements. Though numerous methods exist for calibrating cantilever spring constants, the accuracy of these methods can be limited by both the physical models themselves as well as uncertainties in their experimental implementation. Here we report the results from two of the most common calibration methods, the thermal tune method and the Sader method. These were implemented on a standard AFM system as well as using laser Doppler vibrometry (LDV). Using LDV eliminates some uncertainties associated with optical lever detection on an AFM. It also offers considerably higher signal to noise deflection measurements. We find that AFM and LDV result in similar uncertainty in the calibrated spring constants, about 5%, using either the thermal tune or Sader methods provided that certain limitations of the methods and instrumentation are observed. (c) 2007 American Institute of Physics.