4.6 Article

Trace-level automated mercury speciation analysis

期刊

ANALYTICAL METHODS
卷 3, 期 5, 页码 1143-1148

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c0ay00528b

关键词

-

资金

  1. NIH [P42 ES007373]

向作者/读者索取更多资源

An automated system for methyl Hg analysis by purge and trap gas chromatography (GC) was evaluated, with comparison of several different instrument configurations including chromatography columns (packed column or capillary), detector (atomic fluorescence, AFS, or inductively coupled plasma mass spectrometry, ICP-MS, using quadrupole and sector field ICP-MS instruments). Method detection limits (MDL) of 0.042 pg and 0.030 pg for CH(3)Hg(+) were achieved with the automated Hg analysis system configured with AFS and ICP-MS detection, respectively. Capillary GC with temperature programming was effective in improving resolution and decreasing retention times of heavier Hg species (in this case C(3)H(7)Hg(+)) although carryover between samples was increased. With capillary GC, the MDL for CH(3)Hg(+) was 0.25 pg for AFS detection and 0.060 pg for ICP-MS detection. The automated system was demonstrated to have high throughput (72 samples analyzed in 8 h) requiring considerably less analyst time than the manual method for CH(3)Hg(+) analysis described in EPA 1630.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据