4.6 Article

Development and multi-laboratory verification of US EPA method 538 for the analysis of drinking water contaminants by direct aqueous injection-LC/MS/MS

期刊

ANALYTICAL METHODS
卷 3, 期 7, 页码 1628-1636

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c1ay05134b

关键词

-

向作者/读者索取更多资源

A drinking water method for 11 chemicals, predominantly pesticides, is presented that addresses the occurrence monitoring needs of the U. S. Environmental Protection Agency (EPA) for a future Unregulated Contaminant Monitoring Regulation (UCMR). The method employs direct aqueous injection liquid chromatography/tandem mass spectrometry (DAI-LC/MS/MS). The method uses sodium omadine and ammonium acetate to preserve the drinking water samples for up to 14 days. Mean recoveries of tap water samples fortified with the method analytes at 0.99-4.0 mu g L(-1) (quinoline = 43 mu g L(-1)) are 89.3-106% with relative standard deviations of less than 8%. Single laboratory lowest concentration minimum reporting levels of 0.011-1.5 mu g L(-1) are demonstrated with this methodology. Multi-laboratory data are presented that demonstrate method ruggedness and transferability. The final method meets all of the EPA UCMR survey requirements for sample collection and storage, precision, accuracy, and sensitivity. The final method is expected to be proposed for use under a future UCMR.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据