4.8 Article

Serological Diagnosis of Dengue Infection in Blood Plasma Using Long-Range Surface Plasmon Waveguides

期刊

ANALYTICAL CHEMISTRY
卷 86, 期 3, 页码 1735-1743

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac403539k

关键词

-

资金

  1. Ministry of Higher Education, Malaysia, under High Impact Research Grant [A000007-50001]
  2. Natural Sciences and Engineering Research Council (NSERC) of Canada

向作者/读者索取更多资源

We present a compact, cost-effective, label-free, real-time biosensor based on long-range surface plasmon polariton (LRSPP) gold (Au) waveguides for the detection of dengue-specific immunoglobulin M (IgM) antibody, and we demonstrate detection in actual patient blood plasma samples. Two surface functionalization approaches are proposed and demonstrated: a dengue virus serotype 2 (DENV-2) functionalized surface to capture dengue-specific IgM antibody in blood plasma and the reverse, a blood plasma functionalized surface to capture DENV-2. The results obtained via these two surface functionalization approaches are comparable to, or of greater quality, than those collected by conventional IgM antibody capture enzyme linked immunosorbent assay (MAC-ELISA). Our second functionalization approach was found to minimize nonspecific binding, thus improving the sensitivity and accuracy of the test. We also demonstrate reuse of the biosensors by regenerating the sensing surface down to the virus (or antibody) level or down to the bare Au.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据