4.8 Article

Integrating Biochemiluminescence Detection on Smartphones: Mobile Chemistry Platform for Point-of-Need Analysis

期刊

ANALYTICAL CHEMISTRY
卷 86, 期 15, 页码 7299-7304

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac502137s

关键词

-

向作者/读者索取更多资源

In this paper, we report, for the first time, the use of a smartphone to image and quantify biochemiluminescence coupled biospecific enzymatic reactions to detect analytes in biological fluids. Using low-cost three-dimensional (3D) printing technology, we fabricated a smartphone accessory and a minicartridge for hosting biospecific reactions. As a proof-of-principle, we report two assays: a bioluminescence assay for total bile acids using 3 alpha-hydroxyl steroid dehydrogenase coimmobilized with bacterial luciferase system and a chemiluminescence assay for total cholesterol using cholesterol esterase/cholesterol oxidase coupled with the luminol-H2O2-horseradish peroxidase system. These assays can be performed within 3 min in a very straightforward manner and provided adequate analytical performance for the analysis of total cholesterol in serum (limit of detection (LOD) = 20 mg/dL) and total bile acid in serum and oral fluid (LOD = 0.5 mu mol/L) with a reasonable accuracy and precision. Smartphone-based biochemiluminescence detection could be thus applied to a variety of clinical chemistry assays.