4.8 Article

Part per Trillion Label-Free Electronic Bioanalytical Detection

期刊

ANALYTICAL CHEMISTRY
卷 85, 期 8, 页码 3849-3857

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac302702n

关键词

-

资金

  1. FlexSMELL - Gas Sensors on Flexible Substrates for Wireless Applications [FP7-PITN-GA-2009-238454]
  2. PRIN 09 project Electronic and electrochemical biosensors [2009AZKNJ7]

向作者/读者索取更多资源

A Functional Bio-Interlayer Organic Field-Effect Transistor (FBI-OFET) sensor, embedding a streptavidin protein capturing layer, capable of performing label-free selective electronic detection of biotin at 3 part per trillion (mass fraction) or 15 pM, is proposed here. The response shows a logarithmic dependence spanning over 5 orders of magnitude of analyte concentration. The optimization of the FBI analytical performances is achieved by depositing the capturing layer through a controllable Layer-by-Layer (LbL) assembly, while an easy processable spin-coating deposition is proposed for potential low-cost production of equally highly performing sensors. Furthermore, a Langmuirian adsorption based model allows rationalizing the analyte binding process to the capturing layer. The FBI-OFET device is shown to operate also with an antibody interlayer as well as with an ad hoc designed microfluidic system. These occurrences, along with the proven extremely high sensitivity and selectivity, open to FBI-OFETs consideration as disposable electronic strip-tests for assays in biological fluids requiring very low detection limits.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据