4.8 Article

A Simple Separation Method with a Microfluidic Channel Based on Alternating Current Potential Modulation

期刊

ANALYTICAL CHEMISTRY
卷 84, 期 22, 页码 9738-9744

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac301351y

关键词

-

资金

  1. MEST, South Korea [20100029128]
  2. NRF

向作者/读者索取更多资源

A simple separation and detection system based on an electrochemical potential modulated microchannel (EPMM) device was developed for the first time. The application of alternating current (AC) potential to the microfluidic separation channel walls, which were composed of screen printed carbon electrodes, resulted in the oscillation and fluctuation of analytes and in the formation of a perfect flat flow front. These events resulted in an increase in the effective concentration and in the fine separation of samples. The performance of the EPMM device was examined through the analysis of endocrine disruptors (EDs) and heavy metal ions (HMIs) as model compounds. The analytical parameters that affected the separation and detection of EDs and HMIs were studied in terms of AC amplitude, AC frequency, flow rate, buffer concentration, pH, detection potential, and temperature. The separation efficiency was evaluated through measurements of the theoretical plate number (N), the retention time, and the half-peak width. Linear calibration plots for the detection of EDs and HMIs were obtained between 0.15 and 250.0 nM (detection limit 86.4 +/- 2.9 pM) and between 0.01 and 10.0 nM (detection limit 9.5 +/- 0.3 pM), respectively. The new device was successfully demonstrated with authentic and real samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据