Accuracy of the LifeShirt® (Vivometrics) in the detection of cardiac rhythms

期刊

BIOLOGICAL PSYCHOLOGY
卷 75, 期 3, 页码 300-305

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.biopsycho.2007.04.001

关键词

ambulatory monitoring; heart period; respiratory sinus arrhythmia; baseline; exercise

资金

  1. NIMH NIH HHS [R21 MH067446-03, R01 MH060625-05S1, T32 MH067631-01A1, MH067631, MH067446, R01 MH060625, R21 MH067446, MH 60625, T32 MH067631] Funding Source: Medline

向作者/读者索取更多资源

The use of heart rate measures in research requires accurate detection and timing of beat-to-beat values. Numerous technologies are available to researchers; however, benchmarking of a specific apparatus is seldom conducted. Since heart rate variability provides a portal to the neural regulation of the heart, accurate detection and timing of beat-to-beat values is essential to both basic physiological research and the clinical application of heart rate variability measures. The current study evaluated the accuracy of an ambulatory system, the LifeShirt (R) (Vivometrics), relative to a standard laboratory-based heart rate monitoring equipment (Biopac), during baseline and exercise conditions. LifeShirt (R) performed equivalently to the Biopac during both conditions, experienced few errors of detection, generated similar times between sequential heart periods, and produced similar summary indices of heart rate and heart rate variability. Published by Elsevier B.V.