4.8 Article

Corona Discharge Ion Mobility Spectrometry with Orthogonal Acceleration Time of Flight Mass Spectrometry for Monitoring of Volatile Organic Compounds

期刊

ANALYTICAL CHEMISTRY
卷 84, 期 12, 页码 5327-5334

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac300722s

关键词

-

资金

  1. Slovak research and development 398 agency [LPP-0143-06, APVV-0733-11]
  2. grant agency VEGA [1/0514/12]

向作者/读者索取更多资源

We demonstrate the application of corona discharge ion mobility spectrometry with orthogonal acceleration time of flight mass spectrometry (CD IMS-oaTOF) for volatile organic compounds (VOCs) monitoring. Two-dimensional (2D) IMS-oaTOF spectra of VOCs were recorded in nearly real time. The corona discharge atmospheric pressure chemical ionization (APCI) source was operated in positive mode in nitrogen and air. The CD ion source generates in air H3O+(H2O)(n) and NO+. The NO+ offers additional possibility for selective ionization and for an increase of the sensitivity of monoaromatic compounds. In addition to H3O+(H2O)(n) and NO+, we have carried out ionization of VOCs using acetone as dopant gas ((CH3)(2)COH+). Sixteen model VOCs (tetrahydrofuran, butanol, npropanol, iso-propano, acetone, methanol, ethanol, toluene, benzene, amomnia, dioxan, triethylamine, acetonitrile, formaldehyde, m-xylene, 2,2,2-trifluoroethylamine) were tested using these ionization techniques.