4.5 Article Proceedings Paper

Artificial acrylic finger nails may alter pulse oximetry measurement

期刊

RESUSCITATION
卷 74, 期 1, 页码 75-82

出版社

ELSEVIER IRELAND LTD
DOI: 10.1016/j.resuscitation.2006.11.018

关键词

acrylic artificial finger nails; pulse oximetry; arterial blood gas analysis; measurement error; monitoring

向作者/读者索取更多资源

Introduction: Pulse oximetry is the most common technique to monitor oxygen saturation (SpO(2)) during intensive care therapy. However, intermittent co-oximetry is still the gold standard (SaO(2)). Besides acrylic nails, numerous other factors have been reported to interfere with pulse oximetry. Data of measurements with artificial finger nails are not sufficiently published. Materials and methods: A prospective clinical-experimentat trial in mechanically ventilated and critically ill patients of an ICU was performed. Patients were randomly assigned to either group S (S: Siemens pulse oximeter) or group P (P: Philips pulse oximeter) prior to the measurements. SPO2 was determined in each patient three times alternately in standard (SPO2) and sideways position at the natural nail ((SPO2)-S-N90). For the reference measurements oxygen saturation was measured by means of a haemoximeter (co-oximetry). Thereafter, SPO2 was obtained at the acrylic finger nail in the same way (SPO2)-S-A and (SPO2)-S-A90). Bias was calculated as AS = (SPO2)-S-N-Sa02 and AS = ASpO(2) - SaO2. Accuracy (mean difference) and precision (standard deviation) were used to determine the measurement discrepancy. P < 0.05 was considered significant. Results: Accuracy and precision without acrylic nails applied were comparable to SaO2 in both groups (n.s.). With acrylic nails applied a bias of AS = -1.1 +/- 3.14% for group S (P = 0.00522) and a bias of AS = +0.8 +/- 13.04% for group P was calculated (n.s.). Conclusion: Acrylic finger nails may impair the measurement of oxygen saturation depending on the pulse oximeter used and may cause significant inaccuracy. Hence, removal of artificial acrylic finger nails may be helpful to assure an accurate and precise measurement with pulse oximetry. (C) 2006 Elsevier Ireland Ltd. All. rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据