4.4 Article

Carbonate assignment and calibration in the raman spectrum of apatite

期刊

CALCIFIED TISSUE INTERNATIONAL
卷 81, 期 1, 页码 46-52

出版社

SPRINGER
DOI: 10.1007/s00223-007-9034-0

关键词

carbonated apatite; Raman spectroscopy; bone

资金

  1. NIAMS NIH HHS [R01 AR47969] Funding Source: Medline

向作者/读者索取更多资源

A series of apatites with varying carbonate levels was prepared in order to assign the carbonate bands and calibrate for Raman analysis of natural materials. Overlap of carbonate bands with phosphate peaks was resolved by curve fitting. A peak at 1,071 cm(-1) was assigned to a combination of the carbonate nu(1) mode at 1,070 cm(-1) with a phosphate nu(3) mode at 1,076 cm(-1). In addition, the carbonate nu(4) mode was identified in apatite samples with > 4% carbonate. The carbonate nu(4) bands at 715 and 689 cm(-1) identify the samples as B-type carbonated apatite. The carbonate content of apatite was calibrated to a carbonate Raman band, and the method was used to determine the carbonate content of a sample of bovine cortical bone, 7.7 +/- 0.4%.