4.3 Article

The effect of core framework designs on the fracture loads of all-ceramic fixed partial dentures on posterior implants

期刊

JOURNAL OF ORAL REHABILITATION
卷 34, 期 7, 页码 503-507

出版社

BLACKWELL PUBLISHING
DOI: 10.1111/j.1365-2842.2007.01751.x

关键词

all-ceramic bridge; core framework design; fracture load; zirconium; posterior bridge

向作者/读者索取更多资源

This study evaluated the fracture loads of three-unit all-ceramic bridges on implants; the core frameworks of the bridges were made of zirconium oxide. Three core framework designs were fabricated according to the design of the bars between the retainers: (i) straight, (ii) curved in the occlusal direction, or (iii) curved in the gingival direction. A static load was applied at the centre of the pontic, and the mean initial crack and final fracture loads were measured and compared. The core framework curved in the occlusal direction had the highest final fracture loads; there were significant differences (P < 0.05) in the mean final fracture load, whereas initial fracture load, which fractured of veneered porcelain, did not show significant differences among three designs. The all-ceramic bridges on the implants made with Procera zirconium core frameworks had high final fracture load. The core framework design that curved in the occlusal direction helped the framework withstand the occlusal load, which results in reliable prostheses, especially in the molar region.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据