4.8 Article

Enrichment and Detection of Escherichia coli O157:H7 from Water Samples Using an Antibody Modified Microfluidic Chip

期刊

ANALYTICAL CHEMISTRY
卷 82, 期 7, 页码 2844-2849

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac100323k

关键词

-

资金

  1. National Institutes of Health [1R33-CA09924601]
  2. National Science Foundation [EPS-0346411]
  3. State of Louisiana Board of Regents Support Fund
  4. Texas Parks and Wildlife Department
  5. U.S. Department of Energy
  6. Texas Sea Grant [NA06OAR4170076]

向作者/读者索取更多资源

Low abundant (<100 cells mL(-1)) E. coli O157:H7 cells were isolated and enriched from environmental water samples using a microfluidic chip. The poly(methylmethacrylate), PMMA, chip contained 8 devices, each equipped with 16 curvilinear high aspect ratio channels that were covalently decorated with polyclonal anti-O157 antibodies (pAb) and could search for rare cells through a pAb mediated process. The chip could process independently 8 different samples or one sample using 8 different parallel inputs to increase volume processing throughput. After cell enrichment, cells were released and enumerated using benchtop real-time quantitative polymerase chain reaction (PCR), targeting genes which effectively discriminated the 0157:H7 serotype from other nonpathogenic bacteria. The recovery of target cells from water samples was determined to be similar to 72%, and the limit-of-detection was found to be 6 colony forming units (cfu) using the slt1 gene as a reporter. We subsequently performed analysis of lake and wastewater samples. The simplicity in manufacturing and ease of operation makes this device attractive for the selection of pathogenic species from a variety of water supplies suspected of containing bacterial pathogens at extremely low frequencies.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据