Does the Mass Spectrometer Define the Marker? A Comparison of Global Metabolite Profiling Data Generated Simultaneously via UPLC-MS on Two Different Mass Spectrometers

期刊

ANALYTICAL CHEMISTRY
卷 82, 期 19, 页码 8226-8234

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac1016612

关键词

-

资金

  1. European Reintegration Grant (ERG) [202132]
  2. NIH
  3. National Institute of General Medical Sciences [5R21GM75903]
  4. NATIONAL INSTITUTE OF GENERAL MEDICAL SCIENCES [R21GM075903] Funding Source: NIH RePORTER

向作者/读者索取更多资源

By coupling a single UPLC separation to two different types of mass spectrometer an unbiased comparison of the metabolite profiles produced by each instrument for a set of rat urine samples was obtained. The flow from the UPLC column was split equally and both streams of eluent were simultaneously directed to the inlets of the two mass spectrometers. Mass spectrometry on the eluent was undertaken using a triple quadrupole linear ion trap and a hybrid quadrupole time-of-flight mass spectrometer using both positive and negative ESI. Data from both mass spectrometers were subjected to multivariate statistical analysis, after applying the same data extraction software, and showed the same general pattern of correlation between the samples using both unsupervised and supervised methods of statistical analysis. Based on orthogonal partial least-squares discriminant analysis models a number of ions were recognized as responsible for the separation of the animal groups. From the peaks detected, and denoted as significant by the statistical analysis a number of ions were found to be unique to one data set or the other, a result which may have consequences for biomarker discovery and interlaboratory comparisons. The software package used for data analysis also had an effect on the outcome of the statistical analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据