期刊
ANALYTICAL CHEMISTRY
卷 81, 期 8, 页码 3194-3198出版社
AMER CHEMICAL SOC
DOI: 10.1021/ac802434z
关键词
-
资金
- JSPS [17310087, 20310072]
- Grants-in-Aid for Scientific Research [17310087, 20310072] Funding Source: KAKEN
In about a 3 min period, we have simultaneously separated plasma from human whole blood and metered and diluted the plasma using a microchip with an interchannel microstructure. The plasma separation was based on both cross-flow filtration and sedimentation of red blood cells in the microchannels. Metering and diluting operations of the plasma were based on volume control of liquid in the microchannels by syringe pumps. On this microchip, we produced plasma diluted by a factor of 6 from whole blood containing theophylline and we observed very little hemolysis. It is possible to separate plasma from one or just several drops of whole blood by using this microchip.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据