期刊
BULLETIN OF ENVIRONMENTAL CONTAMINATION AND TOXICOLOGY
卷 79, 期 1, 页码 1-4出版社
SPRINGER
DOI: 10.1007/s00128-007-9172-0
关键词
formaldehyde; headspace analysis; gas chromatography; human urine; O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine
Formaldehyde is a carcinogen to which humans are exposed daily, but few methods are available to quantify formaldehyde in biological samples. We developed a simple, sensitive and rapid technique for the quantification of formaldehyde in urine by derivatization with O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine, using a headspace sampler coupled to a gas chromatograph equipped with an electron capture detector. The detection limit was 1.08 mu g/L. The overall recovery of formaldehyde spiked in urine was 99%. The concentration of formaldehyde in urine obtained from healthy volunteers ranged from 56.85 to 144.57 mu g/L. This method can be used successfully to measure formaldehyde in urine.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。