The influence of patient centering on CT dose and image noise

期刊

MEDICAL PHYSICS
卷 34, 期 7, 页码 3093-3101

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.2748113

关键词

-

向作者/读者索取更多资源

Although x-ray intensity shaping filters (bowtie filters) have been used since the introduction of some of the earliest CT scanner models, the clinical implications on dose and noise are not well understood. To achieve the intended dose and noise advantage requires the patient to be centered in the scan field of view. In this study we explore the implications of patient centering in clinical practice. We scanned various size and shape phantoms on a GE LightSpeed VCT scanner using each available source filter with the phantom centers positioned at 0, 3, and 6 cm below the center of rotation (isocenter). Surface doses were measured along with image noise over a large image region. Regression models of surface dose and noise were generated as a function of phantom size and centering error. Methods were also developed to determine the amount of miscentering using a scout scan projection radiograph (SPR). These models were then used to retrospectively evaluate 273 adult body patients for clinical implications. When miscentered by 3 and 6 cm, the surface dose on a 32 cm CTDI phantom increased by 18% and 41% while image noise also increased by 6% and 22%. The retrospective analysis of adult body scout SPR scans shows that 46% of patients were miscentered in elevation by 20-60 mm with a mean position 23 mm below the center of rotation (isocenter). The analysis indicated a surface dose penalty of up to 140% with a mean dose penalty of 33% assuming that tube current is increased to compensate for the increased noise due to miscentering. Clinical image quality and dose efficiency can be improved on scanners with bowtie filters if care is exercised when positioning patients. Automatically providing patient specific centering and scan parameter selection information can help the technologist improve workflow, achieve more consistent image quality and reduce patient dose. (C) 2007 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据