On-chip single-copy real-time reverse-transcription PCR in isolated picoliter droplets

期刊

ANALYTICAL CHEMISTRY
卷 80, 期 6, 页码 1854-1858

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac800048k

关键词

-

向作者/读者索取更多资源

The first lab-on-chip system for picoliter droplet generation and RNA isolation, followed by reverse transcription, and PCR amplification with real-time fluorescence detection in the trapped droplets has been developed. The system utilized a shearing T-junction in a fused-silica device to generate a stream of monodisperse picoliter-scale droplets that were isolated from the microfluidic channel walls and each other by the oil-phase carrier. An off-chip valving system stopped the droplets on-chip, allowing thermal cycling for reverse transcription and subsequent PCR amplification without droplet motion. This. combination of the established real-time reverse transcription-PCR assay with digital microfluidics is ideal for isolating single-copy RNA and virions from a complex environment and will be useful in viral discovery and gene-profiling applications.