期刊
ANALYTICA CHIMICA ACTA
卷 594, 期 2, 页码 257-264出版社
ELSEVIER
DOI: 10.1016/j.aca.2007.05.029
关键词
aflatoxin M-1; milk; validation; high performance liquid chromatography; fluorescence detection; 2002/657/EC
A high performance liquid chromatographic method with fluorimetric detection for the determination of aflatoxin M, (AFMI) in milk has been optimized and validated according to Commission Decision 2002/657/EC by using the conventional validation approach. The procedure for determining selectivity, recovery, precision, decision limit (CC alpha), detection capability (CC beta) and ruggedness of the method has been reported. The results of the validation process demonstrate the agreement of the method with the provisions of Commission Regulation 401/2006/EC. The mean recovery calculated at three levels of fortification (0.5, 1.0, and 1.5-fold the MRL) was 91 % and the maximum relative standard deviation value forthe within-laboratory reproducibility was 15%. Limit of detection (LOD) and limit of quantitation (LOQ) values were 0.006 Rg kg(-1) and 0.015 mu g kg(-1) while the CC, and CC values were 0.058 Rg kg-1 and 0.065 jig kg(-1), respectively. The relative expanded measurement uncertainty of the method was 7%. The method was not affected by slight variations of some critical factors (ruggedness minor changes) as pre-treatment and clean-up of milk samples, thermal treatment and different storage conditions, as well as by major changes valued in terms of milk produced by different species (buffalo, goat and sheep). The method allowed accurate confirmation analyses of milk samples, resulted positive by the screening method. In fact, the Z-score values attained in a proficiency test round were well below the reference value of 1, proving the excellent laboratory performances. (C) 2007 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据