4.7 Article Proceedings Paper

Simultaneous determination of chromium(III) and cadmium(II) by differential pulse anodic stripping voltammetry on a stannum film electrode

期刊

TALANTA
卷 72, 期 5, 页码 1733-1737

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2007.04.055

关键词

stannum film electrode; chromium(III); cadmium(ll); differential pulse anodic stripping voltammetry

向作者/读者索取更多资源

A stannum film electrode has been developed for the simultaneous determination of trace levels of chromium(III) and cadmium(II) by differential pulse anodic stripping voltammetry (DPASV). The stannum film electrode was generated in situ by depositing simultaneously the stannum film and the metals obtained by reduction of Cd(II) and Cr(III) at -1.4 V on a glassy carbon electrode. Then, the reduced products were oxidized by scanning the potential of the electrode from -1.4 to -0.4 V using DPASV. The electrode exhibited well-defined and separated stripping signals for both metals accompanied with a low background contribution. The possible mechanism of this design was proposed. Under the optimized working conditions, the detection limit was 2.0 and 1.1 mu g l(-1) for Cr(III) and Cd(II) at a deposition time of 3 min. Finally, the stannum film electrode was successfully applied to the determination of Cd(II) in tap water with satisfactory results. (C) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据