4.5 Article

Determination of trace lead in biological and water samples with dispersive liquid-liquid microextraction preconcentration

期刊

ANALYTICAL BIOCHEMISTRY
卷 380, 期 1, 页码 21-25

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.ab.2008.05.008

关键词

dispersive liquid-liquid microextraction; lead; graphite furnace atomic absorption spectrometry; biological and water samples

向作者/读者索取更多资源

A new method for the determination of trace lead was developed by dispersive liquid-liquid microextraction preconcentration and graphite furnace atomic absorption spectrometry, In the proposed approach, 1 -phenyl-3-methyl-4-benzoyl-5-pyrazolone (PMBP) was used as a chelating agent, and carbon tetrachloride and ethanol were selected as extraction and dispersive solvents. Some factors influencing the extraction efficiency of lead and its subsequent determination, including extraction and dispersive solvent type and volume, pH of sample solution, concentration of the chelating agent, and extraction time, were studied and optimized. Under the optimum conditions, the enrichment factor of this method for lead was reached at 78. The detection limit for lead was 39 ng L-1 (3 sigma), and the relative standard deviation (RSD) was 3.2% (n = 7, c = 10 ng mL(-1)). The method was successfully applied to the determination of trace amounts of lead in human urine and water samples. (c) 2008 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据