4.6 Article

Contact-free measurement of cardiac pulse based on the analysis of thermal imagery

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 54, 期 8, 页码 1418-1426

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2007.891930

关键词

adaptive filtering; cardiac pulse; fast Fourier transform (FFT); medical imaging; thermal imaging

向作者/读者索取更多资源

We have developed a novel method to measure human cardiac pulse at a distance. It is based on the information contained in the thermal signal emitted from major superficial vessels. This signal is acquired through a highly sensitive thermal imaging system. Temperature on the vessel is modulated by pulsative blood flow. To compute the frequency of modulation (pulse), we extract a line-based region along the vessel. Then, we apply fast Fourier transform (FFT) to individual points along this line of interest to capitalize on the pulse's thermal propagation effect. Finally, we use an adaptive estimation function on the average FFT outcome to quantify the pulse. We have carried out experiments on a data set of 34 subjects and compared the pulse computed from our thermal signal analysis method to concomitant ground-truth measurements obtained through a standard contact sensor (piezo-electric transducer). The performance of the new method ranges from 88.52% to 90.33% depending on the clarity of the vessel's thermal imprint. To the best of our knowledge, it is the first time that cardiac pulse has been measured several feet away from a subject with passive means.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据