4.5 Article

Feasibility of two multidimensional techniques, heart-cut MDGC and GC x GC, for the separation of PCBs and PBDEs

期刊

JOURNAL OF SEPARATION SCIENCE
卷 30, 期 12, 页码 1920-1929

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.200700045

关键词

comprehensive two dimensional GC; heart-cut MDGC; PBDEs; PCBs; separation

向作者/读者索取更多资源

The feasibility of heart-cut MDGC-ECD and GC x GC-mu ECD for the separation of polybrominated diphenyl ethers (PBDEs) and polychlorinated biphenyls (PCBs) has been evaluated. Among the different column combinations tested, the best results were obtained when using either a DB-5 type or an HT-8 column as the first dimension combined with a mid-polar stationary phase (50% phenyl) as the second dimension. The coelutions between the target PCBs and PBDEs in the first dimension were satisfactorily resolved once transferred to the second column. Repeatability and intermediate precision were satisfactory in both systems, considering retention times in both dimensions. Nevertheless, peak area/volume precision was better for heart-cut MDGC since software for the data analysis in the case of GC x GC was still under development. The need for multiple injections of the same extract on the heart-cut MDGC system was the main drawback since the analysis time it increased. GC x GC becomes then a valuable alternative able to achieve the same performance in one injection. However, the GC x GC data analysis is still very time-consuming and needs further development. The applicability of both separation techniques was shown using a human breast milk sample with low concetrations of PCBs (0.2-227 pg/g fresh weight) and PBDEs (1.2-41 pg/g fresh weight).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据