4.7 Article

Silicone discs as disposable enrichment probes for gas chromatography-mass spectrometry determination of UV filters in water samples

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 400, 期 2, 页码 603-611

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-011-4763-3

关键词

Sample preparation; Sorptive extraction; Silicone sorbents; UV filters; Water analysis

资金

  1. Spanish Government
  2. EU [CTQ2009-08377, DE2009-0020]
  3. FPU

向作者/读者索取更多资源

This work describes an effective, low solvent consumption and affordable sample preparation approach for the determination of eight UV filters in surface and wastewater samples. It involves sorptive extraction of target analytes in a disposable, technical grade silicone disc (5 mm diameter x 0.6 mm thickness) followed by organic solvent desorption, large volume injection (LVI), and gas chromatography-mass spectrometry determination. Final working conditions involved overnight extraction of 100-mL samples, containing 10% of methanol, followed by analytes desorption with 0.2 mL of ethyl acetate. The method provides linear responses between the limits of quantification (from 0.003 to 0.040 ng mL(-1)) and 10 ng mL(-1), an intra-day precision below 13%, and low matrix effects for surface, swimming pool, and treated sewage water samples. Moreover, the extraction yields provided by silicone discs were in excellent agreement with those achieved using polydimethylsiloxane-covered stir bars. Several UV filters were found in surface and sewage water samples, with the maximum concentrations corresponding to octocrylene.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据