JOURNAL OF APPLIED CRYSTALLOGRAPHY
卷 40, 期 -, 页码 791-795出版社
BLACKWELL PUBLISHING
DOI: 10.1107/S0021889807022248
关键词
-
A novel apparatus has been developed that enables the simultaneous determination of the absorption factor during measurement of small-angle X-ray scattering (SAXS) intensities of a sample. It was designed especially for the use of relatively low-energy X-rays at SAXS beamlines of synchrotron facilities. The X-ray intensity of transmittance is measured by a silicon PIN photodiode, which is implanted in a direct beamstop set in a vacuum chamber. Since the assembly transmits an attenuated direct beam to a detector during the scattering measurement, a zero-angle position can be monitored without additional operation. It was confirmed that the linearity between the signal from the photodiode and the intensity of X-rays is good and the photodiode is applicable for the desired purpose. For a performance test, the absorption factors of a supercritical fluid were measured with a wide density range.