ULTRAMICROSCOPY
卷 107, 期 8, 页码 663-668出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.ultramic.2007.01.002
关键词
TEM sample preparation; ion milling; temperature
类别
The actual temperature rise was measured during ion-milling process used in the transmission electron microscopy (TEM) sample preparation. Special probes were fabricated for the measurements, one with shielded, floating thermocouple mounted onto a 3 mm grid to compute the thermal load at the sample, and the other, a bare probe with a polymer coating to measure the maximum temperature attained. The temperature measured in the most commonly used ion-milling system reached up to 295 degrees C when the typical milling conditions, 6keV ion-energy and an incident angle of 80 degrees, were used. Based on the temperature profiles that were obtained by the shielded probe, two unknown parameters, the amount of heat deposited by the energetic ions/neutrals to the sample and the thermal conductivities between the materials, were estimated and used to compute the temperature rise in commonly adopted materials. The calculation showed that the temperature of the glass sample reached more than 300 degrees C under typical ion-milling conditions. The calculated value was confirmed with the experimental result of the crystallization of an amorphous Si on the glass under the typical ion-milling condition, which gave the same extent as annealing at 350 degrees C. (C) 2007 Elsevier B.V. All rights reserved.