A further method for chloride analysis of cement and cementitious materials - ICP-OES

期刊

CEMENT AND CONCRETE RESEARCH
卷 37, 期 8, 页码 1172-1175

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.cemconres.2007.05.006

关键词

chloride analysis; OPC; cementitious materials; ICP-OES; potentiometric analysis

向作者/读者索取更多资源

This paper describes the novel application of inductively coupled plasma-optical emission spectroscopy (ICP-OES) for the analysis of the chloride content in various South African cements and cementitious materials. Samples were introduced into the plasma as slurries, with and without 5% (v/v) nitric acid. They were agitated in an ultrasonic bath for 10 min to fully disperse and stabilize the slurries. Calibration was performed by using aqueous solutions of reference materials. Measurements were made on a simultaneous SpectroCIROS(CCD) ICP-OES with side on plasma. The accuracy of the chloride determination was verified by the analysis of certified reference material. The findings of this investigation are compared with results obtained for the chloride content by the standard EN196-2 indirect Volhard titration and potentiometry with an ion-selective electrode (ISE), previously described as the preferred method. The results obtained with potentiometry and ICP-OES of various commercially available cements were comparable. (c) 2007 Elsevier Ltd. All fights reserved.