期刊
ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 392, 期 4, 页码 681-698出版社
SPRINGER HEIDELBERG
DOI: 10.1007/s00216-008-2292-5
关键词
glucocorticoids; anti-oestrogens; designer steroids; stimulants; diuretics; beta-adrenergic drugs; LC-ESI-MS/MS; anti-doping analysis
This paper presents a general screening method, based on liquid chromatography/mass spectrometry (LC/MS), for the simultaneous detection in human urine of 72 xenobiotics (21 diuretics, 16 synthetic glucocorticoids, 17 beta-adrenergic drugs, 10 stimulants, 5 anti-oestrogens and 3 anabolic steroids), excreted free or as glucuro-conjugates in urine. Although the method has been specifically designed and evaluated in view of its potential application to anti-doping analyses, it can also be effective in other areas of analytical toxicology. Sample preparation was based on two liquid/liquid separation steps (performed at alkaline and at acid pH, respectively) of hydrolyzed human urine, and then an assay by LC/MS-MS in positive and negative ionization mode using an electrospray ionization source (ESI) and multiple reaction monitoring (MRM) as the acquisition mode. The overall time needed for an LC run was less than 15 minutes. All compounds showed good reproducibility in terms of both the retention times (CV%< 1) and the relative abundances of the diagnostic transitions (CV%< 10). The limits of detection (LOD) were in the range of 1-50 ng/mL for glucocorticoids, anti-oestrogens and steroids, and 50-500 ng/mL for diuretics, beta-adrenergic drugs and stimulants, thus satisfying the minimum required performance limits (MRPL) set by the World Anti-Doping Agency (WADA) for the accredited anti-doping laboratories.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐