4.7 Article Proceedings Paper

Titania sol-gel-derived tyrosinase-based amperometric biosensor for determination of phenolic compounds in water samples. Examination of interference effects

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 391, 期 4, 页码 1275-1281

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-007-1798-6

关键词

tyrosinase; amperometric biosensor; phenols; Plackett-Burman design

向作者/读者索取更多资源

For detection of phenolic compounds in environmental water samples we propose an amperometric biosensor based on tyrosinase immobilized in titania sol gel. The analytical characteristics toward catechol, p-cresol, phenol, p-chlorophenol, and p-methylcatechol were determined. The linear range for catechol determination was 2.2 x 10(-7) - 1.3 x 10(-5) mol L-1 with a limit of detection of 9 x 10(-8) mol L-1 and sensitivity 2.0 x 10(3) mA mol(-1) L. The influence of sample matrix components on the electrode response was studied according to Plackett-Burman experimental design. The potential interferents Mg2+, Ca2+, HCO3-, SO42- , and Cl-, which are usually encountered in waters, were taken into account in the examination. Cu2+ was also taken into account, because CuSO4 is sometimes added to a water sample, as a preservative, before determination of phenolic compounds. It was found that among the ions tested only Mg2+ and Ca2+ did not directly affect the electrode response. The developed biosensor was used for determination of catechol in spring and surface water samples using the standard addition method.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据