4.7 Article

Headspace sorptive extraction using silicone tubes for the determination of chlorobenzenes in water

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 393, 期 2, 页码 767-775

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-008-2455-4

关键词

Silicone tube; Headspace; Chlorobenzenes; Water; Thermodesorption-gas chromatography-mass spectrometry

向作者/读者索取更多资源

A new approach for headspace sorptive extraction is presented and demonstrated for the determination of 12 chlorobenzenes in water samples. It consists of a silicone tube (15-mm length) arranged around a stainless steel rod. This device is fixed on a septum cap and exposed to the headspace of 50 mL of a salt-saturated water sample. After extraction (60-min optimized extraction time), thermodesorption is carried out by direct insertion of the silicone tube into the thermodesorption-gas chromatography-mass spectrometry system. Desorption of the analytes is performed at 250 degrees C for 5 min with a gas flow of 50 mL/min. Repeatability (relative standard deviation 5-10%), extraction yields (9-46%), enrichment factors (129-657), and detection limits (0.002-0.012 mu g/L) were determined and four real water samples were analyzed with the headspace tube extraction. The results were verified by standard addition. A comparison of headspace tube extraction with other headspace enrichment techniques underlined the high extraction capacity of the proposed method. A big advantage of tube extraction is the low cost of the silicone material. The tubes can be discarded after single use, avoiding carryover problems and cross-contamination.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据