A novel gas sampling introduction interface for fast analysis of volatile organic compounds using radiofrequency pulsed glow discharge time of flight mass spectrometry

期刊

ANALYTICA CHIMICA ACTA
卷 1038, 期 -, 页码 59-66

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2018.08.013

关键词

Glow discharge; Mass spectrometry; Volatile organic compounds (VOCs) analysis; Gas chromatography; Direct injection; Air analysis

资金

  1. Ministerio de Educacion y Cultura [MINECO-13-CTQ2013-49032-C2-2-R, MINECO-17-CTQ2016-77887-C2-1-R]

向作者/读者索取更多资源

An improved gas sample introduction interface is developed and characterized for gas chromatography coupling and for direct injection of volatile organic compounds (VOCs), in a pulsed glow discharge (pulsed-GD) ion source coupled to a time of flight mass spectrometer (TOFMS) that is typically used for direct solid analysis. The novel interface allows the introduction of the analytes in the flowing afterglow region of the GD (a few mm away from the negative glow region) to reduce plasma quenching effects. Analyte ion signals are acquired in the temporal afterglow region, where low fragmentation of the molecular species is produced, providing useful qualitative and quantitative molecular information (e.g. molecular ion). Analytical capabilities of the pulsed-GD ion source with the novel gas sampling interface provides improved performance compared to previous designs. In particular, limits of detection for the analysis of VOCs in air were below (better) that legally established limits according to Directive 2008/50/EC of the European Parliament. (C) 2018 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据