4.6 Article

Production of radioactive quality assurance phantoms using a standard inkjet printer

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 52, 期 15, 页码 N329-N337

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/52/15/N02

关键词

-

向作者/读者索取更多资源

This note proposes the use of a standard inkjet printer to produce radioactive Tc-99m phantoms that can be used for routine quality control of gamma cameras. The amount of activity that will be deposited on paper per unit area was predicted and compared to the measured activity. The uniformity of the printouts was compared to the uniformity obtained with a standard Co-57 flood source. A scintillation detector connected to a multi-channel analyzer was used to evaluate the uniformity of the printout independently from the gamma camera. Joining two A4 size printed phantoms to create larger sources was evaluated. The spatial resolution obtained with printed sources was compared to that obtained using standard line source techniques. The results indicated that the uniformity of the printed phantoms compared well with those obtained with the 57Co flood source (integral uniformity 2.29% (printed source) and 2.10% (57Co flood source)). There was no difference in the resolution measurements obtained with the printed sources and those obtained with the standard methods. This study demonstrates that affordable phantoms can easily be created to evaluate system uniformity and resolution in any department where a standard PC and inkjet printer are available.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据