期刊
ANALYTICA CHIMICA ACTA
卷 706, 期 1, 页码 84-88出版社
ELSEVIER
DOI: 10.1016/j.aca.2011.07.046
关键词
Inductively coupled plasma mass spectrometry; Polyatomic interferences; Matrix effects; Dynamic reaction cell; Trace elements; Seawater
A method for the determination of ultratrace amounts of Cr, Fe, Mn, Pb and Zn in seawater has been developed. It combined the low-blank magnesium hydroxide coprecipitation procedure with quadrupole inductively coupled plasma mass spectrometry and used the dynamic reaction cell technique to resolve the polyatomic interferences arising from the residual matrix, the solvent and plasma gases. Detection limits (3 sigma(B), n = 10) for Cr, Fe, Mn, Pb and Zn were 0.02, 0.10, 0.01, 0.002 and 0.19 nM, respectively, using 50 mL of seawater sample. The accuracy of the analytical procedure was verified by the analysis of the seawater reference materials CASS-4, NASS-5, SAFe D2 and SAFe S. The analytical precision ranged from 3% to 16% (n = 6), with a sample throughput of about 6 samples h(-1). (C) 2011 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐