4.7 Article

A urea biosensor based on pH-sensitive Sm2TiO5 electrolyte-insulator-semiconductor

期刊

ANALYTICA CHIMICA ACTA
卷 669, 期 1-2, 页码 68-74

出版社

ELSEVIER
DOI: 10.1016/j.aca.2010.04.045

关键词

Urea; Biosensor; pH-sensitive; Sm2TiO5; Electrolyte-insulator-semiconductor

资金

  1. National Science Council, Taiwan, Republic of China [NSC-98-2221-E-182-056-MY3]

向作者/读者索取更多资源

A urea biosensor based on pH-sensitive Sm2TiO5 electrolyte-insulator-semiconductor (EIS) has been described. We used X-ray diffraction, Auger electron spectroscopy, and atomic force microscopy to investigate the structural and morphological features of high-k Sm2TiO5 sensing membranes that had been subjected to annealing at different temperatures. The EIS device incorporating a high-k Sm2TiO5 sensing film that had been annealed at 900 degrees C exhibited good sensing characteristics, including a high sensitivity of 60.5 mV/pH (in solutions from pH 2 to 12), a small hysteresis voltage of 2.72 mV (in the pH loop 7 -> 4 -> 7 -> 10 -> 7), and a low drift rate of 1.15 mV h(-1) (in the buffer solution at pH 7). The Sm2TiO5 EIS device also showed a high selective response towards H+. This improvement can be attributed to the small number of crystal defects and the large surface roughness. in addition, the urea biosensor based on pH-sensitive EIS incorporating a Sm2TiO5 sensing membrane annealed at 900 degrees C allowed the potentiometric analysis of urea, at concentrations ranging from 0.1 to 32 mM, with a sensitivity of 72.85 mV/purea. (C) 2010 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者